问题:SIWAREX FTA用于贸易交接时,如果采用WinCC flexible设计触摸屏程序,必须考虑哪些内容?
解答: 对于每一个版本的WinCC flexible都会单独提供一个SecureOCX控件。根据用户手册第8.6节的内容进行安装,SecureOCX的版本必须在称重模块的DR39中设置。.
目前可以提供下列版本的SecureOCX控件:
WinCC flexible | Secure OCX版本 | 状态 |
WinCC flexible 2005 SP1 HF6 or 7 | V2.1 | 可以提供 |
WinCC flexible 2007 | V2.2 | 可以提供 |
WinCC flexible 2008 without SP | V2.3 | 可以提供 |
WinCC flexible 2008 SP1 | V2.4 | 可以提供 |
WinCC flexible 2008 SP2 | V2.5 | 可以提供 |
WinCC flexible 2008 SP3 | V2.6 | 可以提供 |
After updating the secure OCX to a new version, the "software ID" of data record 39 inside the SIWAREX module must be changed to the new version number too! This procedure is only possible with an opened write-protection switch - already calibrated scales must be recalibrated!